N2 - Objective: Commercial cochlear-implant (CI) telehealth is manufacturer-specific and generally does not replicate standard sound-field tests. This study developed and tested a manufacturer-independent CI telehealth system designed to replace a standard clinic visit. Design: The system comprised a laptop, a sound card, headphones (placed over the CI sound processor), manufacturer’s programming software and cables, and custom software for aided speech-recognition and warble-tone detection-threshold assessment. Phase 1 compared telehealth-system aided assessment to standard sound-field testing. In Phase 2, participants completed at-home remote audiological follow-up visits. Study Sample: N = 15 (Phase 1) and N = 13 (Phase 2) adult CI users. Results: Mean absolute differences between telehealth-system and sound-field warble-tone detection thresholds were 5 dB, with 96% of 0.5–4 kHz telehealth-system thresholds within the sound-field 95% confidence interval (95ci). Consonant-nucleus-consonant word and AzBio sentence identification in quiet showed no significant differences between modes, with high intraclass correlations (ICC = 0.94–0.97) and 80-93% of telehealth-system scores within the sound-field 95ci. Bamford-Kowal-Bench sentence-in-noise scores showed larger deviations. Telehealth-survey responses indicated high patient satisfaction, improved convenience and access to care, and only minor challenges. Conclusions: Combining a manufacturer-independent headphones-based alternative to sound-field testing with manufacturer’s programming software in a remotely controlled laptop system replicates sound-field outcome scores and produces high patient satisfaction.
AB - Objective: Commercial cochlear-implant (CI) telehealth is manufacturer-specific and generally does not replicate standard sound-field tests. This study developed and tested a manufacturer-independent CI telehealth system designed to replace a standard clinic visit. Design: The system comprised a laptop, a sound card, headphones (placed over the CI sound processor), manufacturer’s programming software and cables, and custom software for aided speech-recognition and warble-tone detection-threshold assessment. Phase 1 compared telehealth-system aided assessment to standard sound-field testing. In Phase 2, participants completed at-home remote audiological follow-up visits. Study Sample: N = 15 (Phase 1) and N = 13 (Phase 2) adult CI users. Results: Mean absolute differences between telehealth-system and sound-field warble-tone detection thresholds were 5 dB, with 96% of 0.5–4 kHz telehealth-system thresholds within the sound-field 95% confidence interval (95ci). Consonant-nucleus-consonant word and AzBio sentence identification in quiet showed no significant differences between modes, with high intraclass correlations (ICC = 0.94–0.97) and 80-93% of telehealth-system scores within the sound-field 95ci. Bamford-Kowal-Bench sentence-in-noise scores showed larger deviations. Telehealth-survey responses indicated high patient satisfaction, improved convenience and access to care, and only minor challenges. Conclusions: Combining a manufacturer-independent headphones-based alternative to sound-field testing with manufacturer’s programming software in a remotely controlled laptop system replicates sound-field outcome scores and produces high patient satisfaction.
KW - cochlear implant
KW - sound-field testing
KW - speech recognition testing
KW - streaming
KW - Teleaudiology
KW - telemedicine
UR - https://www.scopus.com/pages/publications/105053004931
U2 - 10.1080/14992027.2026.2731750
DO - 10.1080/14992027.2026.2731750
M3 - Article
C2 - 42850951
AN - SCOPUS:105053004931
SN - 1499-2027
JO - International Journal of Audiology
JF - International Journal of Audiology
ER -