Bradycardia-Dependent Loss of Atrial Capture After Single-Chamber Leadless Atrial Pacemaker Implantation

  • Matthew A. Soderstrom*
  • , Ariana Ramirez
  • , John Henry Dean
  • , Mark Haigney
  • , Eric Martinez
  • , Cory Madigan
  • *Corresponding author for this work

Research output: Contribution to journal › Article › peer-review

Abstract

Background: Bradycardia-dependent pacing threshold elevation is a rarely reported cause of early postimplant loss of capture, with only 1 prior case reported in an atrial leadless pacemaker. Case Summary: 88-year-old woman with sinus pauses underwent Aveir AR implantation. On postoperative day 1, intermittent loss of atrial capture occurred at the programmed lower rate limit. Interrogation revealed stable impedance, current of injury, and sensing. Threshold testing revealed marked rate-dependence: 2.75 V/0.4 milliseconds at 60 beats/min vs 0.75 V/0.4 milliseconds at 90 beats/min. Fluoroscopy confirmed device stability, and output was increased conservatively. Discussion: Proposed mechanisms of this phenomenon include rate-dependent variation in the electrode-tissue interface and phase 4 block from diastolic depolarization in injured myocardium. When early intermittent capture loss occurs with stable device parameters, threshold testing across pacing rates should be performed to assess for bradycardia-dependent threshold elevation. Take-Home Message: Bradycardia-dependent capture threshold elevation is a possible etiology of early postimplant loss of atrial capture after Aveir AR implantation.

Original languageEnglish
Article number110232
JournalJACC: Case Reports
Volume31
Issue number43
DOIs
StatePublished - 28 Oct 2026

Keywords

  • bradycardia
  • cardiac pacemaker
  • electrophysiology

Fingerprint

Dive into the research topics of 'Bradycardia-Dependent Loss of Atrial Capture After Single-Chamber Leadless Atrial Pacemaker Implantation'. Together they form a unique fingerprint.

Cite this