Comparative evaluation of intravenous therapy regulating devices

Research output: Contribution to journalArticlepeer-review

2 Scopus citations

Abstract

To determine the patient's benefit from intravenous therapy with infusion pumps, two types of volumetric (ml/hr) intravenous infusion pumps were evaluated in relation to patient complications, cost, and nursing time required for maintaining an intravenous infusion. Based on the complications associated with the infusion devices evaluated, our study has found that the optimal infusion pump may generate a pressure range between 4 and 10 psi. This pressure range may promote the maintenance of intravenous fluid flow, prevent hazards, and minimize cost to the patient. Further research should be conducted to evaluate the optimal pressure infusion device and other infusion devices in relation to patient benefit.

Original languageEnglish
Pages (from-to)262-267
Number of pages6
JournalHeart and Lung
Volume15
Issue number3
StatePublished - 1986