TY - JOUR
T1 - Effect of Varying Fixture Width on Stress and Strain Distribution Associated with an Implant Stack System
AU - Mahon, Joseph M.
AU - Norling, Barry K.
AU - Phoenix, Rodney D.
PY - 2000
Y1 - 2000
N2 - The purpose of this investigation was to evaluate the dissipation of a force applied to an assembled stack of implant components. The stack consisted of a 10-mm threaded implant, a screw-retained abutment and a screw-retained gold crown. The dissipation of force was analyzed in relation to varying the implant diameter with and without a concomitant change in abutment diameter. Two experimental groups were evaluated. The first group consisted of 25 titanium screw-form implants (Implant Innovations, Inc.). These implants measured 10 mm in length and 3.25 mm, 3.75 mm, 4.0 mm, 5.0 mm, and 6.0 mm in diameter. The second group included 15 titanium screw-form implants (Nobel Biocare, Inc.) measuring 10 mm in length and 3.75 mm, 4.0 mm, and 5.0 mm in diameter. All implants were embedded in standardized photoelastic resin blocks. Points of interest were marked on each block using standardized templates to ensure consistency. Implants were restored using system-specific conical abutments and standardized single-unit restorations. A strain gauge was affixed to each abut ment, and an eccentric load of 176 N was applied to the restoration. Periimplant stresses were measured using photoelastic analysis. Abutment strain was determined using an electronic strain indicator. Data were collated and compared using ANOVA and the Duncan multiple range statistical tests. When stress was analyzed at points on the resin-implant interface or a fixed distance from the interface, stress tended to decrease from the 5-mm-wide implant to the 6-mm-wide implant. Stress in relation to the 3.25-mm, 3.75-mm, and 4.0-mm implant was not as well defined, indicating the possibility that some deformation of implants was occurring. Increased abutment width resulted in decreased abutment strain. Therefore, using a wider abutment may be helpful in preventing preload reduction in clinical applications. This may reduce the incidence of loosening and fracture of abutment and restoration screws.
AB - The purpose of this investigation was to evaluate the dissipation of a force applied to an assembled stack of implant components. The stack consisted of a 10-mm threaded implant, a screw-retained abutment and a screw-retained gold crown. The dissipation of force was analyzed in relation to varying the implant diameter with and without a concomitant change in abutment diameter. Two experimental groups were evaluated. The first group consisted of 25 titanium screw-form implants (Implant Innovations, Inc.). These implants measured 10 mm in length and 3.25 mm, 3.75 mm, 4.0 mm, 5.0 mm, and 6.0 mm in diameter. The second group included 15 titanium screw-form implants (Nobel Biocare, Inc.) measuring 10 mm in length and 3.75 mm, 4.0 mm, and 5.0 mm in diameter. All implants were embedded in standardized photoelastic resin blocks. Points of interest were marked on each block using standardized templates to ensure consistency. Implants were restored using system-specific conical abutments and standardized single-unit restorations. A strain gauge was affixed to each abut ment, and an eccentric load of 176 N was applied to the restoration. Periimplant stresses were measured using photoelastic analysis. Abutment strain was determined using an electronic strain indicator. Data were collated and compared using ANOVA and the Duncan multiple range statistical tests. When stress was analyzed at points on the resin-implant interface or a fixed distance from the interface, stress tended to decrease from the 5-mm-wide implant to the 6-mm-wide implant. Stress in relation to the 3.25-mm, 3.75-mm, and 4.0-mm implant was not as well defined, indicating the possibility that some deformation of implants was occurring. Increased abutment width resulted in decreased abutment strain. Therefore, using a wider abutment may be helpful in preventing preload reduction in clinical applications. This may reduce the incidence of loosening and fracture of abutment and restoration screws.
KW - Abutment width
KW - Implant width
KW - Photoelastic analysis
KW - Strain gauge
UR - http://www.scopus.com/inward/record.url?scp=0034570063&partnerID=8YFLogxK
U2 - 10.1097/00008505-200009040-00006
DO - 10.1097/00008505-200009040-00006
M3 - Article
C2 - 11307553
AN - SCOPUS:0034570063
SN - 1056-6163
VL - 9
SP - 310
EP - 318
JO - Implant Dentistry
JF - Implant Dentistry
IS - 4
ER -