TY - JOUR
T1 - Electrical impedance scanning for the early detection of breast cancer in young women
T2 - Preliminary results of a multicenter prospective clinical trial
AU - Stojadinovic, Alexander
AU - Nissan, Aviram
AU - Gallimidi, Zahava
AU - Lenington, Sarah
AU - Logan, Wende
AU - Zuley, Margarita
AU - Yeshaya, Arieh
AU - Shimonov, Mordechai
AU - Melloul, Moshe
AU - Fields, Scott
AU - Allweis, Tanir
AU - Ginor, Ron
AU - Gur, David
AU - Shriver, Craig D.
PY - 2005/4/20
Y1 - 2005/4/20
N2 - Purpose: To evaluate the feasibility and patient satisfaction with electrical impedance scanning (EIS) for early detection of breast cancer in young women. Methods: Women undergoing screening clinical breast examination, imaging, or biopsy were eligible for EIS examination with T-Scan 2000ED (Mirabel Medical Systems, Austin, TX). Multiple logistic regression analysis evaluated the association between clinical variables and EIS performance. Patients completed a screening EIS satisfaction questionnaire (1 = least satisfied to 5 = most satisfied). Results: Twenty-nine cancers were identified among 1,103 women. Sixty-six percent (19 of 29) of cancers were nonpalpable and 55% (16 of 29) were in women age < 50 years. EIS sensitivity and specificity in women younger than 40 years was 50% and 90%, respectively. Exogenous estrogen use (P < .001) and menopausal status (P = .007) correlated significantly with EIS performance. False-positive rates were increased in postmenopausal women and those taking exogenous hormones. No correlation was evident between EIS performance and family history, prior breast cancer, breast density, or palpability. EIS-positive women younger than age 40 were 4.5 times more likely to have breast carcinoma than were women randomly selected from the general population. Patients were highly satisfied with the comfort, speed, and reporting of EIS screening (mean score, 4.8). Conclusion: EIS seems promising for early detection of breast cancer, and identification of young women at increased risk for having the disease at time of screening. Positive EIS-associated breast cancer risk compares favorably with relative risks of conditions commonly used to justify early breast cancer screening. Patients are satisfied with a screening paradigm involving breast EIS.
AB - Purpose: To evaluate the feasibility and patient satisfaction with electrical impedance scanning (EIS) for early detection of breast cancer in young women. Methods: Women undergoing screening clinical breast examination, imaging, or biopsy were eligible for EIS examination with T-Scan 2000ED (Mirabel Medical Systems, Austin, TX). Multiple logistic regression analysis evaluated the association between clinical variables and EIS performance. Patients completed a screening EIS satisfaction questionnaire (1 = least satisfied to 5 = most satisfied). Results: Twenty-nine cancers were identified among 1,103 women. Sixty-six percent (19 of 29) of cancers were nonpalpable and 55% (16 of 29) were in women age < 50 years. EIS sensitivity and specificity in women younger than 40 years was 50% and 90%, respectively. Exogenous estrogen use (P < .001) and menopausal status (P = .007) correlated significantly with EIS performance. False-positive rates were increased in postmenopausal women and those taking exogenous hormones. No correlation was evident between EIS performance and family history, prior breast cancer, breast density, or palpability. EIS-positive women younger than age 40 were 4.5 times more likely to have breast carcinoma than were women randomly selected from the general population. Patients were highly satisfied with the comfort, speed, and reporting of EIS screening (mean score, 4.8). Conclusion: EIS seems promising for early detection of breast cancer, and identification of young women at increased risk for having the disease at time of screening. Positive EIS-associated breast cancer risk compares favorably with relative risks of conditions commonly used to justify early breast cancer screening. Patients are satisfied with a screening paradigm involving breast EIS.
UR - http://www.scopus.com/inward/record.url?scp=21044437997&partnerID=8YFLogxK
U2 - 10.1200/JCO.2005.06.155
DO - 10.1200/JCO.2005.06.155
M3 - Article
C2 - 15837985
AN - SCOPUS:21044437997
SN - 0732-183X
VL - 23
SP - 2703
EP - 2715
JO - Journal of Clinical Oncology
JF - Journal of Clinical Oncology
IS - 12
ER -