Field trial of the carestart biosensor analyzer for the determination of glucose-6-phosphate dehydrogenase activity in Haiti

Thomas A. Weppelmann, Michael E. Von Fricken, Tara D. Wilfong, Elisa Aguenza, Taina T. Philippe, Bernard A. Okech*

*Corresponding author for this work

Research output: Contribution to journalArticlepeer-review

16 Scopus citations

Abstract

Throughout many developing and tropical countries around the world, malaria remains a significant threat to human health. One barrier to malaria elimination is the ability to safely administer primaquine chemotherapy for the radical cure of malaria infections in populations with a high prevalence of glucose-6-phosphate dehydrogenase (G6PD) deficiency. In the current study, a field trial of the world's first quantitative, point-of-care assay for measuring G6PD activity was conducted in Haiti. The performance of the CareStart Biosensor Analyzer was compared with the gold standard spectrophotometric assay and genotyping of the G6PD allele in schoolchildren (N = 343) from the Ouest Department of Haiti. In this population, 19.5% of participants (67/343) had some form of G6PD deficiency (< 60% residual activity) and 9.9% (34/343) had moderate-to-severe G6PDdeficiency (<30%residual activity). Overall, 18.95% of participants had the presence of the A-allele (65/343) with 7.87% (27/343) considered at high risk for drug-induced hemolysis (hemizygous males and homozygous females). Compared with the spectrophotometric assay, the sensitivity and specificity to determine participants with < 60% residual activity were 53.7% and 94.6%, respectively; for participants with 30% residual activity, the sensitivity and specificity were 5.9% and 99.7%, respectively. The biosensor overestimated the activity in deficient individuals and underestimated it in participants with normal G6PD activity, indicating the potential for a systematic measurement error. Thus, we suggest that the current version of the biosensor lacks adequate sensitivity and should be improved prior to its use as a point-of-care diagnostic for G6PD deficiency.

Original languageEnglish
Pages (from-to)1262-1270
Number of pages9
JournalAmerican Journal of Tropical Medicine and Hygiene
Volume97
Issue number4
DOIs
StatePublished - 2017
Externally publishedYes

Fingerprint

Dive into the research topics of 'Field trial of the carestart biosensor analyzer for the determination of glucose-6-phosphate dehydrogenase activity in Haiti'. Together they form a unique fingerprint.

Cite this