T1 - Intramedullary Unicortical Button and All-Suture Anchors Provide Similar Maximum Strength for Onlay Distal Biceps Tendon Repair
AU - Colantonio, Donald F.
AU - Le, Anthony H.
AU - Keeling, Laura E.
AU - Slaven, Sean E.
AU - Vippa, Tarun K.
AU - Helgeson, Melvin D.
AU - Chang, Edward S.
N1 - Publisher Copyright:
© 2021
PY - 2022/2
Y1 - 2022/2
N2 - Purpose: To evaluate the biomechanical profile of onlay distal biceps repair with an intramedullary unicortical button versus all-suture anchors under cyclic loading and maximal load to failure. Methods: Twenty paired fresh-frozen human cadaveric elbows were randomized to onlay distal biceps repair with either a single intramedullary button or with two 1.35-mm all-suture anchors. A 1.3-mm high tensile strength tape was used in a Krackow stitch to suture the tendons in both groups. Specimens and repair constructs were loaded for 3,000 cycles and then loaded to failure. Maximum load to failure, mode of failure, and construct elongation were recorded. Results: Mean (± standard deviation) maximum load to failure for the unicortical intramedullary button and all-suture anchor repairs were 503.23 ± 141.77 N and 537.33 ± 262.13 N (P = .696), respectively. Mean maximum displacement after 3,000 cycles (± standard deviation) was 4.17 ± 2.05 mm in the button group and 2.06 ± 1.05 mm in the suture anchor group (P = .014). Mode of failure in the button group was suture tape rupture in 7 specimens, failure at the tendon–suture interface in 2 specimens, and button pullout in 1 specimen. Anchor pullout was the mode of failure in all suture anchor specimens. There were no tendon ruptures or radial tuberosity fractures in either group. Conclusions: This study demonstrates that onlay distal biceps repair with 2 all-suture anchors has similar maximum strength to repair with an intramedullary button and that both are viable options for fixation. Clinical Relevance: All-suture anchors and unicortical intramedullary button have similar maximum strength at time zero. Both constructs provide suitable fixation for onlay distal biceps repair.
AB - Purpose: To evaluate the biomechanical profile of onlay distal biceps repair with an intramedullary unicortical button versus all-suture anchors under cyclic loading and maximal load to failure. Methods: Twenty paired fresh-frozen human cadaveric elbows were randomized to onlay distal biceps repair with either a single intramedullary button or with two 1.35-mm all-suture anchors. A 1.3-mm high tensile strength tape was used in a Krackow stitch to suture the tendons in both groups. Specimens and repair constructs were loaded for 3,000 cycles and then loaded to failure. Maximum load to failure, mode of failure, and construct elongation were recorded. Results: Mean (± standard deviation) maximum load to failure for the unicortical intramedullary button and all-suture anchor repairs were 503.23 ± 141.77 N and 537.33 ± 262.13 N (P = .696), respectively. Mean maximum displacement after 3,000 cycles (± standard deviation) was 4.17 ± 2.05 mm in the button group and 2.06 ± 1.05 mm in the suture anchor group (P = .014). Mode of failure in the button group was suture tape rupture in 7 specimens, failure at the tendon–suture interface in 2 specimens, and button pullout in 1 specimen. Anchor pullout was the mode of failure in all suture anchor specimens. There were no tendon ruptures or radial tuberosity fractures in either group. Conclusions: This study demonstrates that onlay distal biceps repair with 2 all-suture anchors has similar maximum strength to repair with an intramedullary button and that both are viable options for fixation. Clinical Relevance: All-suture anchors and unicortical intramedullary button have similar maximum strength at time zero. Both constructs provide suitable fixation for onlay distal biceps repair.
UR - http://www.scopus.com/inward/record.url?scp=85113194772&partnerID=8YFLogxK
U2 - 10.1016/j.arthro.2021.06.036
DO - 10.1016/j.arthro.2021.06.036
M3 - Article
C2 - 34332050
AN - SCOPUS:85113194772
SN - 0749-8063
VL - 38
SP - 287
EP - 294
JO - Arthroscopy - Journal of Arthroscopic and Related Surgery
JF - Arthroscopy - Journal of Arthroscopic and Related Surgery
IS - 2
ER -