T1 - Migration of Aortic Occlusion Balloons in an in vitro model of the human circulation
AU - Borger van der Burg, B. L.S.
AU - Van Schaik, J.
AU - Brouwers, J. J.W.M.
AU - Wong, C. Y.
AU - Rasmussen, T. E.
AU - Hamming, J. F.
AU - Hoencamp, R.
N1 - Publisher Copyright:
© 2018 Elsevier Ltd
PY - 2019/2
Y1 - 2019/2
N2 - Background: Aortic Occlusion Balloons (AOB) are used for hemorrhage control in hemodynamically unstable patients. Stability of an AOB is essential for reliable aortic occlusion. The primary aim of this study is to determine whether different types of AOB migrate after total, intermittent or partial occlusion in a porcine aorta positioned in an in vitro model. Materials and methods: A porcine thoracic aortic section was positioned in a model of the human circulation. Primary and secondary migration was tested in Cook Coda™ 2–9.0-35-120-32 and 2-10-35-140-46, Cook Medical, USA; Rescue balloon™ Tokai RB-167080-E, Tokai Medical Products, Japan; Reliant™ AB46, Medtronic, USA; Russian prototype AOB; ER-REBOA™, Prytime Medical Devices, USA; LeMaitre™ 28 and 45 Aortic Occlusion Catheter, LeMaitre Vascular, USA. These AOB were tested in hypotensive, normotensive and hypertensive scenarios. Migration in total occlusion, intermittent occlusion and partial occlusion was recorded for all AOB. Results: Limited primary migration occurred in all AOB after total occlusion. The Cook Coda™ 2–9.0-35-120-32 balloon showed maximal migration in 1 test cycle. No migration occurred during intermittent occlusion. Kinking occurs in various degrees but does not seem to prevent a successful occlusion of the aorta. No migration occurred during partial occlusion except in the Russian prototype AOB. In a partial occlusion scenario, distal perfusion occurred only with 5 ml remaining in all balloon types. Conclusions: All AOB were successful in full aortic occlusion. Limited primary migration occurred in all AOB after total occlusion only the Cook Coda™ 2–9.0-35-120-32 balloon showed maximal migration once. No migration occurred during intermittent occlusion, during partial occlusion only the Russian prototype AOB migrated. Stiffness and size of the catheter are important factors in preventing migration and kinking.
AB - Background: Aortic Occlusion Balloons (AOB) are used for hemorrhage control in hemodynamically unstable patients. Stability of an AOB is essential for reliable aortic occlusion. The primary aim of this study is to determine whether different types of AOB migrate after total, intermittent or partial occlusion in a porcine aorta positioned in an in vitro model. Materials and methods: A porcine thoracic aortic section was positioned in a model of the human circulation. Primary and secondary migration was tested in Cook Coda™ 2–9.0-35-120-32 and 2-10-35-140-46, Cook Medical, USA; Rescue balloon™ Tokai RB-167080-E, Tokai Medical Products, Japan; Reliant™ AB46, Medtronic, USA; Russian prototype AOB; ER-REBOA™, Prytime Medical Devices, USA; LeMaitre™ 28 and 45 Aortic Occlusion Catheter, LeMaitre Vascular, USA. These AOB were tested in hypotensive, normotensive and hypertensive scenarios. Migration in total occlusion, intermittent occlusion and partial occlusion was recorded for all AOB. Results: Limited primary migration occurred in all AOB after total occlusion. The Cook Coda™ 2–9.0-35-120-32 balloon showed maximal migration in 1 test cycle. No migration occurred during intermittent occlusion. Kinking occurs in various degrees but does not seem to prevent a successful occlusion of the aorta. No migration occurred during partial occlusion except in the Russian prototype AOB. In a partial occlusion scenario, distal perfusion occurred only with 5 ml remaining in all balloon types. Conclusions: All AOB were successful in full aortic occlusion. Limited primary migration occurred in all AOB after total occlusion only the Cook Coda™ 2–9.0-35-120-32 balloon showed maximal migration once. No migration occurred during intermittent occlusion, during partial occlusion only the Russian prototype AOB migrated. Stiffness and size of the catheter are important factors in preventing migration and kinking.
KW - Aortic occlusion balloon
KW - Migration
KW - Model
KW - REBOA
UR - http://www.scopus.com/inward/record.url?scp=85059016259&partnerID=8YFLogxK
U2 - 10.1016/j.injury.2018.12.026
DO - 10.1016/j.injury.2018.12.026
M3 - Article
C2 - 30594315
AN - SCOPUS:85059016259
SN - 0020-1383
VL - 50
SP - 286
EP - 291
JO - Injury
JF - Injury
IS - 2
ER -