TY - JOUR
T1 - Modular extracorporeal life support
T2 - Effects of ultrafiltrate recirculation on the performance of an extracorporeal carbon dioxide removal device
AU - Scaravilli, Vittorio
AU - Kreyer, Stefan
AU - Linden, Katharina
AU - Belenkiy, Slava
AU - Jordan, Bryan
AU - Pesenti, Antonio
AU - Zanella, Alberto
AU - Chung, Kevin
AU - Cannon, Jeremy
AU - Cancio, Leopoldo C.
AU - Batchinsky, Andriy I.
PY - 2014
Y1 - 2014
N2 - The combination of extracorporeal carbon dioxide removal (ECCO2R) and hemofiltration is a possible therapeutic strategy for patients needing both lung and renal support. We tested the effects of the recirculation of ultrafiltrate on membrane lung (ML) CO2 removal (VCO2ML). Three conscious, spontaneously breathing sheep were connected to a commercially produced ECCO2R device (Hemolung; Alung Technologies, Pittsburgh, PA) with a blood flow of 250 ml/min and a gas flow of 10 L/min. A hemofilter (NxStage, NxStage Medical, Lawrence, MA) was interposed in series after the ML. Ultrafiltrate flow was generated and recirculated before the ML. We tested four ultrafiltrate flows (0, 50, 100, and 150 ml/min) for 25 min each, eight times per animal, resulting in 24 randomized test repetitions. We recorded VCO2ML, hemodynamics and ventilatory variables, and natural lung CO2 transfer (VCO2NL) and collected arterial and circuitry blood samples. VCO2ML was unchanged by application of ultrafiltrate recirculation (40.5 ± 4.0, 39.7 ± 4.2, 39.8 ± 4.2, and 39.2 ± 4.1 ml/min, respectively, at ultrafiltrate flow of 0, 50, 100, and 150 ml/min). Minute ventilation, respiratory rate, VCO2NL, and arterial blood analyses were not affected by ultrafiltrate recirculation. In the tested configuration, ultrafiltrate recirculation did not affect VCO2ML. This modular technology may provide a suitable platform for coupling CO2 removal with various forms of blood purification.
AB - The combination of extracorporeal carbon dioxide removal (ECCO2R) and hemofiltration is a possible therapeutic strategy for patients needing both lung and renal support. We tested the effects of the recirculation of ultrafiltrate on membrane lung (ML) CO2 removal (VCO2ML). Three conscious, spontaneously breathing sheep were connected to a commercially produced ECCO2R device (Hemolung; Alung Technologies, Pittsburgh, PA) with a blood flow of 250 ml/min and a gas flow of 10 L/min. A hemofilter (NxStage, NxStage Medical, Lawrence, MA) was interposed in series after the ML. Ultrafiltrate flow was generated and recirculated before the ML. We tested four ultrafiltrate flows (0, 50, 100, and 150 ml/min) for 25 min each, eight times per animal, resulting in 24 randomized test repetitions. We recorded VCO2ML, hemodynamics and ventilatory variables, and natural lung CO2 transfer (VCO2NL) and collected arterial and circuitry blood samples. VCO2ML was unchanged by application of ultrafiltrate recirculation (40.5 ± 4.0, 39.7 ± 4.2, 39.8 ± 4.2, and 39.2 ± 4.1 ml/min, respectively, at ultrafiltrate flow of 0, 50, 100, and 150 ml/min). Minute ventilation, respiratory rate, VCO2NL, and arterial blood analyses were not affected by ultrafiltrate recirculation. In the tested configuration, ultrafiltrate recirculation did not affect VCO2ML. This modular technology may provide a suitable platform for coupling CO2 removal with various forms of blood purification.
KW - Extracorporeal carbon dioxide removal
KW - hemofiltration
KW - recirculation
KW - respiratory dialysis
UR - http://www.scopus.com/inward/record.url?scp=84900395390&partnerID=8YFLogxK
U2 - 10.1097/MAT.0000000000000070
DO - 10.1097/MAT.0000000000000070
M3 - Article
C2 - 24614356
AN - SCOPUS:84900395390
SN - 1058-2916
VL - 60
SP - 335
EP - 341
JO - ASAIO Journal
JF - ASAIO Journal
IS - 3
ER -