Renal angioplasty and stent placement with distal protection: Preliminary experience with the FilterWire EX

Klaus D. Hagspiel*, James R. Stone, Daniel A. Leung

*Corresponding author for this work

Research output: Contribution to journalArticlepeer-review

44 Scopus citations

Abstract

Percutaneous transluminal renal angioplasty and stent placement with use of a coronary protection device was performed in a total of four patients with hypertension (n = 4) and/or renal insufficiency (n = 3) referred for revascularization of five renal arteries. Renal revascularization was successful in all five renal arteries (100%), but renal revascularization under protection by the FilterWire EX was achieved in only three of five renal arteries (60%). In one of these three, only a suboptimal seal was achieved between the vessel wall and the filter basket. Nevertheless, use of the device was safe and fibrin and/or cholesterol fragments were retrieved from three renal arteries. The FilterWire EX needs to be modified for the renal circulation to achieve the full theoretical advantages of these systems in this vascular bed.

Original languageEnglish
Pages (from-to)125-131
Number of pages7
JournalJournal of Vascular and Interventional Radiology
Volume16
Issue number1
DOIs
StatePublished - Jan 2005
Externally publishedYes

Fingerprint

Dive into the research topics of 'Renal angioplasty and stent placement with distal protection: Preliminary experience with the FilterWire EX'. Together they form a unique fingerprint.

Cite this